BENLEDNINGSMÄTNING VID ANVÄNDANDE AV BYGEL OCH SOFTBAND En jämförande studie av luft-ben-gap och självupplevd komfort vid olika anläggningstryck
BONE CONDUCTION TESTING WITH THE USE OF HEADBAND AND SOFTBAND A comparative study of air-bone-gaps and self-estimated comfort at different static force levels
Abstract
Syfte:
Syftet med denna studie har varit att undersöka om det ses någon skillnad i mätresultat mellan bygel och softband samt om det finns någon skillnad i upplevd komfort.
Metod:
Sjutton personer deltog i studien. Fastställning av bentrösklar utfördes med bentelefon B-71 vid användande av bygel P-3333 och två olika softband. Skillnad i luft-ben-gap mellan bygel och softband har undersökts för frekvenserna 0,5 till 4 kHz. Vid 2 kHz undersöktes också luft-ben-gap samt komfort vid olika tryck (3 N, 4 N, 5,4 N). Gradering av komfort gjordes för alla tryck på en modifierad Borg-skala.
Resultat:
Det sågs en statistisk signifikant skillnad i mätresultat mellan softband och bygel vid alla frekvenser förutom 0,5 kHz. Softbandet fick sämre svar än luftledning vid alla frekvenser utom 1 kHz. Den största skillnaden i luft-ben-gap sågs vid 4 kHz med ett medelvärde på 5,27 dB till bygelns fördel. Deltagarna upplevde en större komfort vid användande av softband och lägre tryck gav högre komfortgradering.
Konklusion:
Enligt denna studie kan det konstateras att det finns en statistisk signifikant skillnad i mätresultaten mellan bygel och softband men endast vid 4 kHz är denna skillnad kliniskt relevant. Ingen signifikant skillnad vid 2 kHz benledningströskel fanns vid de olika trycken hos softbandet. Ett softband kan vara ett bra verktyg till testbatteriet exempelvis när benledaren konstant flyttar på sig eller anomalier som försvårar placering av bygel. Ett softband framtaget för B-71 bör användas kliniskt.
Degree
Student essay
Other description
Purpose:
The purpose of this study was to investigate if there were any differences between bone conduction threshold testing results using headband and softband. Self-estimated comfort results were also investigated.
Method:
Seventeen individuals participated in the study. Determination of bone conduction threshold was tested with bone vibrator B-71 with the use of headband P-3333 and two different softbands. The differences between air-bone-gaps for headband and softband were investigated for the frequencies 0,5 to 4 kHz. At 2 kHz, the air-bone gap and self-estimated comfort were investigated at different static force levels (3 N, 4 N, 5.4 N). Valuation of self-estimated comfort was done for all static force levels using a modified Borg-scale.
Result:
Statistically significant differences were found for the test results between headband and softband at all frequencies except for 0.5 kHz. The bone conduction threshold using the softband had worse results than air conduction threshold at all frequencies except 1kHz. The biggest air-bone-gap difference was seen at 4 kHz with an average difference on 5.27 dB in favor of the headband. The participants experienced greater comfort using the softband and lowering the static force levels increased comfort ratings.
Conclusion:
According to this study there is a statistically significant difference in the test results between the headband and the softband, although only at 4 kHz it can be considered clinically relevant. No significant difference was found for bone conduction thresholds between the static force levels tested with the softband at 2 kHz. Softband can be a good complement to the audiological test battery in situations where headband is subpar due to anomalies of the skull. A softband specifically designed for the B-71 bone vibrator is recommended to be used clinically.
Collections
View/ Open
Date
2017-06-29Author
Vässmar, Johan
Karlsson, Marcus
Keywords
Benledningsmätning
Bygel
Softband
Luft-ben-gap
Komfort
Bone Conduction
Headband
Air-Bone-Gap
Comfort
Series/Report no.
2017-010